Zhifeng Sun,Kaixiang Yang,Hongtao Chen,Tao Sui,Lei Yang,Dawei Ge,Jian Tang,Xiaojian Cao.Journal of Biomedical Research,2018,32(2):123-129
A novel entry point for pedicle screw placement in the thoracic spine
Received:March 15, 2016  Revised:June 07, 2016
DOI10.7555/JBR.31.20160037
Keywordsentry point, screw placement, thoracic spine, biomechanical stability
Grant Program
                       
AuthorInstitution
Zhifeng Sun Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Kaixiang Yang Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Hongtao Chen Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Tao Sui Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Lei Yang Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Dawei Ge Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Jian Tang Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Xiaojian Cao Department of Orthopaedics, the First Affiliated Hospital of Nanjing Medical University, Nanjing, Jiangsu , China
Hits: 21
Download times: 20
      
Abstract
      This study was aimed to introduce a novel entry point for pedicle screw fixation in the thoracic spine and compare it with the traditional entry point. A novel entry point was found with the aim of improving accuracy, safety and stability of pedicle screw technique based on anatomical structures of the spine. A total of 76 pieces of normal thoracic CT images at the transverse plane and the thoracic pedicle anatomy of 6 cadaveric specimens were recruited. Transverse pedicle angle (TPA), screw length, screw placement accuracy rate and axial pullout strength of the two different entry point groups were compared. There were significant differences in the TPA, screw length, and the screw placement accuracy rate between the two groups (P < 0.05). The maximum axial pullout strength of the novel entry point group was slightly larger than that of the traditional group. However, the difference was not significant (P > 0.05). The novel entry point significantly improved the accuracy, stability and safety of pedicle screw placement. With reference to the advantages above, the new entry point can be used for spinal internal fixations in the thoracic spine.